Air and gas analysis
Measured / controlled parameters
Mass concentration of aerosol particles PM1, PM2.5 and PM10
Measurement / analysis range
0.001 – 150 mg/m³
Measurement accuracy / precision
±0.1% of reading or ±0.001 mg/m³ (whichever is greater); zero stability ±0.002 mg/m³ per 24 hours
Temperature measurement range
—
Temperature compensation
—
Analysis method / operating principle
90° light-scattering laser photometer
Measurable concentration range of gases / substances
0.001 – 150 mg/m³ (aerosol particles)
Response time / analysis time
1 second (minimum logging interval)
Sensor type
Optical laser light-scattering sensor
Number of measurement channels
Simultaneous measurement of PM1, PM2.5 and PM10
Availability of automatic mode / programmable cycles
Automatic 24/7 continuous operation with data logging
Material of the body / working part
Outdoor enclosure designed for environmental protection
Housing protection class (IP)
Weatherproof outdoor enclosure (IP rating specified by manufacturer for outdoor use)
Interfaces and outputs
USB, Ethernet, analog output, alarm outputs
Software / PC compatibility
Compatible with PC software for data download and analysis
Memory / data storage availability
Built-in data logging system
Display / indication type
Built-in display for operational parameters
Power supply
AC mains power supply
Operating conditions
Temperature 0 – 50 °C; humidity 0 – 95% non-condensing; designed for outdoor installation
Dimensions
Depending on configuration
Weight
Depending on configuration
Scope of supply
Monitor, outdoor enclosure, sampling system, power cable
Certificates and compliance with standards
CE
Additional functions
Continuous data logging, integration into data transmission systems, alarm outputs